GD-261D Fully-automatic Pensky-Martens Closed Cup Flash Point Analyzer
-
USD51-1 Piece/Pieces
-
USD1≥2 Piece/Pieces
- Payment Type:
- L/C, T/T, Western Union
- Incoterm:
- FOB, CFR, CIF
- Min. Order:
- 1 Piece/Pieces
- Min. Order:
- 1 Piece/Pieces
- Delivery Time:
- 7 Days
- Transportation:
- Ocean, Land, Air
Your message must be between 20 to 2000 characters
Contact NowPlace of Origin: | ChongQing,China |
---|---|
Productivity: | 50piece/Month |
Supply Ability: | 50Piece/Month |
Payment Type: | L/C,T/T,Western Union |
Incoterm: | FOB,CFR,CIF |
Transportation: | Ocean,Land,Air |
Introduction:
» GD-261D Pensky-Martens Flash Point Apparatus is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester.
» GD-261D Pensky-Martens Flash Point Apparatus is used to make determination of the closed cup flash point of the petroleum products.
Main Technical Features:
♣ It adopts LCD screen to display. Full English man-machine dialog interface. It can preset the parameters of expected flash point,sample mark number, atmospheric pressure, test date,etc.It has menu to prompt and input function of guide type.
Main Technical Specifications:
1 |
Power supply
|
AC(220±10%)V,50Hz |
2 | Temperature measurement | Range: room temperature~300°C |
Repeatability:≤3°C | ||
Reproducibility:≤6°C | ||
Accuracy:0.1°C | ||
3 | Heating rate | Procedure A:(5~6)°C/min |
Procedure B:(1~1.6)°C/min | ||
Automatic control and manually adjustable. | ||
4 | Stirring rate | (1) Procedure A: (90~120)RPM |
(2) Procedure B: (250±10)RPM | ||
Automatic control and manually adjustable. | ||
5 | Igniting mode | Electric ignition. Gas flame diameter:3.2mm~4.8mm |
6 | Working condition | ambient temperature:(10~40)°C |
relative humidity:≤80% | ||
7 | Maximum power consumption | 500W |
8 | Dimension | 520mm×360mm×310mm |
Related Keywords