Rbot ASTM D2272 Testing Equipment for Turbine Oils
- Payment Type:
- T/T
Quantity:
Your message must be between 20 to 2000 characters
Contact NowBasic Info
Basic Info
Place of Origin: | China |
---|---|
Payment Type: | T/T |
Product Description
Product Description
I. Summary
GD-0193 RBOT ASTM D2272 Testing Equipment for Turbine Oils is designed and made as per industrial standard of People's Republic of China SH/T0193-92 Lubricating oils-Determination of oxidation stability-Rotating pressure vessel method and ASTM D2272.
GD-0193 RBOT ASTM D2272 Testing Equipment for Turbine Oils is used to determine the oxidation stability of steam turbine with the same composition ( base stock and additives). Also can be used to determine new mineral insulating oil containing 2, 6-BHT.
II. Main technical specifications and parameters
1. Power supply: AC220V± 10%
2. Heating tube power: 2500W
3. Range of pressure sensor: (0~1.6)MPa, accuracy: ± 2‰
4. Temperature controlling range of oil bath: Room temperature~200.0° C, adjustable continuously. 150.0° C in common use.
5. Temperature controlling accuracy: ± 0.1° C
6. Test sample: Two-bomb design, can do two samples at one time. Convenient to do parallel test.
7. Rotation speed: (100± 5)r/min
8. Included angle between oxygen bomb and water level: 30°
9. Cubage of oil bath: 30L
GD-0193 RBOT ASTM D2272 Testing Equipment for Turbine Oils is designed and made as per industrial standard of People's Republic of China SH/T0193-92 Lubricating oils-Determination of oxidation stability-Rotating pressure vessel method and ASTM D2272.
GD-0193 RBOT ASTM D2272 Testing Equipment for Turbine Oils is used to determine the oxidation stability of steam turbine with the same composition ( base stock and additives). Also can be used to determine new mineral insulating oil containing 2, 6-BHT.
II. Main technical specifications and parameters
1. Power supply: AC220V± 10%
2. Heating tube power: 2500W
3. Range of pressure sensor: (0~1.6)MPa, accuracy: ± 2‰
4. Temperature controlling range of oil bath: Room temperature~200.0° C, adjustable continuously. 150.0° C in common use.
5. Temperature controlling accuracy: ± 0.1° C
6. Test sample: Two-bomb design, can do two samples at one time. Convenient to do parallel test.
7. Rotation speed: (100± 5)r/min
8. Included angle between oxygen bomb and water level: 30°
9. Cubage of oil bath: 30L
Related Keywords